Maintenance & Calibration
93
9 Press the
DISPLAY SELECT 1
key to step to the next calibration step. When the
uppermost display line indicates
Ext. Cal
.
Completed
, this external calibration
procedure is completed.
10. When the procedure is complete, then the uppermost display line indicates
Ext.
Cal. Completed. Y
ou may select to store the calibration data collected by
pressing any of the
DISPLAY SELECT
keys or the
ENTER
key. If any input failed
calibration, then its calibration data will not be updated by this action. To abort
the calibration without storing any data collected, then press the
CLEAR
key at
any time during the procedure.
Performing a Partial Calibration
In certain conditions, it may be desired to only partially perform a calibration
procedure. This is achieved by following the same procedure outlined above, but
skipping any input or output level by either not pressing the respective input’s
DISPLAY SELECT
key or by pressing the
DISPLAY SELECT 1
key immediately during
each step.
Performing a Calibration Check Only
The standard external calibration procedure shown above may also be used for a
calibration check only.
Ø
To perform a calibration check:
1, Note the percentage deviations displayed
2. When all steps have been completed press the
CLEAR
key, to discard the
calibration data collected.