Calibration
16
CALIBRATION
To perform a calibration, proceed as follows:
1.
Press the CAL enable button to enter the calibration mode.
2.
Press the UP-DOWN button to select the range to be calibrated.
3.
Short the probes together and wait a minimum period of time as
specified in Reference Table 1. Then press the SHIFT and
TRIG keys at the same time.
4.
Connect the Resistor Standard to the meter in a 4-wire
configuration, refer to Figure 9. Wait for a period of time as
specified in Reference Table 1, for the XT560 to stabilize. Then
press the SHIFT and HOLD keys at the same time.
5.
Repeat this procedure adjusting the XT560 milliohmmeter for
each range: 300mΩ range, 3Ω range, … and finally the 30MΩ
range.