- 19 -
5.11 Technical data
Thermostat
Type
Power rat-
ing Watt
Case Dimen-
sions
LxWxH (mm)
Unit fuse in the
mains cable
Weight
ca. Kg
Dimensions
heating bath
LxWxT (mm)
Heating blocks capacity
HP-1V
300
382 x 212 x 140
T 4,0 A
6
140 x 99 x 75
1 HB-HP1
HP-2V
600
440 x 253 x 140
T 6,3 A
8
198 x 140 x 75
2 HB-HP1 oder 1 HB-HP-2
HP-1F
300
382 x 212 x 140
T 4,0 A
6
-
permanently installed
HP-2F
600
440 x 253 x 140
T 6,3 A
8
-
permanently installed
EC-1V
300
382 x 212 x 140
T 4,0 A
6
148 x 107 x 75
1 HB-EC1
EC-2V
600
440 x 253 x 140
T 6,3 A
8
214 x 107 x 75
2 HB-EC1 oder 1 HB-EC2
BIO-1V
300
335 x 193 140
T 4,0 A
5
140 x 99 x 46
1 HB-BIO1
BIO-2V
600
395 x 233 140
T 6,3 A
7
198 x 140 x 46
2 HB-BIO1 oder 1 HB-BIO2
BIO-1F
300
335 x 193 140
T 4,0 A
5
-
permanently installed
BIO-2F
600
395 x 233 140
T 6,3 A
7
-
permanently installed
LS-1V
300
335 x 193 140
T 4,0 A
5
148 x 107 x 60
1 HB-LS1 or 2 HB-LS1H
LS-2V
600
395 x 233 140
T 6,3 A
7
214 x 107 x 60
1 HB-LS2 or 2 HB-LS1 oder 4
HB-LS1H
QS-1V
91 x 91 x 65
1 HB-QS
QS-1F
Technical modifications reserved!
Voltage
230 V/50,60 Hz
Max. Set-points acc. to units:
HP, EC, QS-Thermostats
130°C oder 210°C
BIO, LS-Thermostats
130°C
Protective system acc. to IEC 529
IP 20
EMV-indications concerning fail-
ure classes
EN 50081-1
EN 50082-2
Length of Mains cable
2m
Controller fuse
T 2,0 A
6. Putting into operation / Operation
6.1 Sample vessels
Note that the diameter of the used test-tubes is less smaller than the diameter of the hole
thus there is only a very narrow gap between the aluminium wall and the tube. This precon-
dition provides a good heat transfer from the heating block into the sample and the maxi-
mum performance of the tempering system can be guaranteed.
Note the limited temperature stability of plastic vessels.
If hazardous vapours are to be expected the thermostat has to be placed in a fume
hood.