Vibration Measurement Instruments

Modern machines often run at high speed and bearing-supports 

are fl exible, foundations should/can be evaluated as fl exible, even 

when it is not mounted on rubber or springs.

The ISO 10816-3 standard allows for slightly higher limits when 

a foundation is considered fl exible than when if it is rigid.  

A conclusion from this is also that a resonance condition in 

principle is not allowed or at least must be avoided at operating 

speeds. In practice this also includes the double speed as well as 

any other natural excitation frequency such as blade passage etc.

A great advantage with proper vibration measurements and the 

use of vibration standards is that you can judge the future 

maintenance cost reliably already at fi rst start-up. If you fi nd 

levels above 3 mm/s RMS, you can be rather sure that the machine 

will cause increased activities in maintenance. 

The specifi c cost and action is of course individual to the 

machine design.

The next logical step is therefore to apply more advanced 

analyzers like VIBER X2™, X3™ or X5™ to learn the frequency 

behind the vibration and thus the exact mechanical fault.  

The practice of this is beyond the scope of this manual.

 
    

 

 

Manual Viber X1_121002.indd   11

10/3/12   9:26 AM

Group 3

:

Pumps with multivane impeller and with separate driver with 

rated power above 15kW.

Group 4:

Pumps with multivane impeller and with integrated driver 

with rated power above 15kW.

Modern machines often run at high speed and bearing-

supports are flexible, foundations should/can be evaluated as 

flexible, even when it is not mounted on rubber or springs.

The ISO 10816-3 standard allows for slightly higher limits 

when a foundation is considered flexible than  rigid. A conclu

-

sion from this is a resonance condition should not be allowed 

or at least must be avoided at operating speeds. In practice, 

this also includes the double speed as well as any other natural 

excitation frequency such as blade passage.

A great advantage with proper vibration measurements and 

the use of vibration standards is that you can judge the future 

maintenance cost reliably at start-up. If you find levels above 

3 mm/s RMS, it’s a risk that the machine will have higher 

maintenance cost. The specific cost and action is specific to 

the machine design.

The next logical step if you want to improve the accuracy of 

measurements is to use more advanced analyzers like VIBER 

X2™, X3™ or X5™ to learn and detect the frequency behind 

the vibration and thus the exact mechanical fault. The practice 

of this is beyond the scope of this manual.

Summary of Contents for Viber X

Page 1: ...VIBER X1 Manual ER X1 Manual Ver 1 0 VIBER X...

Page 2: ...or double safety This is especially important when the machine is remote controlled VMI International AB takes no responsibility for any accidents on people and machines VMI International AB and our a...

Page 3: ...cause of the large amount of information As a result of this we might change and correct these items in later issues without further notice Also changes in the VIBER X1 equipment may take place that a...

Page 4: ...domestic standardisation authority This standard is an upgrade of older standards that has been in use for several decades and has a worldwide acceptance as a good judgement for continuous and long l...

Page 5: ...easurement of bearing condition will start The instrument measures instantaneously a bearing condition value in the range between 500 Hz to 20 000 Hz To go back to vibration measurements press the sym...

Page 6: ...Instruments Functions Press the start key and the instrument starts to measure vibration default is 2 maximum frequency range the diodes indicates low and high pass lters setting The instrument will b...

Page 7: ...ments on vertical or horizontal machinery should follow the three perpendicular axes of true vertical How to make good measurements The sensitivity direction of the transducer coincides with the cente...

Page 8: ...ote that in some cases the mass of the transducer may in uence the reading A good rule is to consider readings on surfaces that are lower in mass than 10 times the mass of the transducer All normal me...

Page 9: ...vibration Notice that a reduction of the unbalance from 4 mm s to 1 mm s will reduce the total value from 6 7 mm s to 5 5 mm s 4 4 2 2 5 5 6 7mm How to interpret vibration measurements A user with no...

Page 10: ...hine s stiff body resonance related to the basic running speed of the machine For example a machine supported by rubber or springs have a resonance at low running speeds The machine starts vibrate at...

Page 11: ...s is beyond the scope of this manual Group 3 Pumps with multivane impeller and with separate driver with rated power above 15kW Group 4 Pumps with multivane impeller and with integrated driver with ra...

Page 12: ...and common findings The following is an extraction of part of the old standard ISO 2372 class 4 large machines on flexible foundations with some common findings added Use this simplified list as a fi...

Page 13: ...Press the bearing symbol The instrument measures the bearing condition value Bearing condition value is the RMS value of all high frequency vibrations in the range of 500 Hz to 16 000 Hz Bearing cond...

Page 14: ...lue would also include vibrations from unbalance misalignment etc and not only from bearing vibrations and would therefore be dif cult to interpret If the selected frequency band only includes very hi...

Page 15: ...he frequency range 500 16 000 Hz this can give a high bearing condition value without the bearing being damaged A high bearing condition value can also be indicated if the bearing is poorly lubricated...

Page 16: ...d or overloaded Technical data VIBER X Accelerometer Standard nom 100 mV g Adjusted to the instrument Input amplitude range Max 199 9 mm s Max 19 99 gBC Dynamic range 60 dB 159 16 Hz Frequency range 1...

Page 17: ...E 3 pc of AAA Rechargeable VIBRATION TRANSDUCER The accelerometer VMI 192 has a sensitivity 100mV g 10 MAGNET SUPPORT Length 20mm Diameter 15mm Magnet force about 14kg Note You must avoid to use other...

Page 18: ...ion of conformity VMI declares that the VIBER X is manufactured in conformity with national and international regulations The system complies with and is tested according to following requirements EMC...

Page 19: ...er the warranty period rst contact your distributor you purchased the unit from Warranty disclaimer VMI warrants the products to be free from defects in material and workmanship under normal use and s...

Page 20: ...VMI International AB Sweden www vmiab com...

Reviews: