1.Safety Information

 

 

same voltage as the LO Input terminal, unless a current protection fuse is open.   

Sense Terminals (HI Sense and LO Sense) Measurement Limits 

The HI and LO sense terminals are used for four-wire resistance measurements.   

The Measurement Limit from HI Sense to LO Input is 200 Vpk. 

The Measurement Limit from HI Sense to LO Sense is 200 Vpk. 

The Measurement Limit from LO Sense to LO Input is 2 Vpk. 

Note

: The 200 Vpk limit on the sense terminals is the Measurement Limit. Operational 

voltages in resistance measurements are much lower – up to ± 3 V in normal operation. 

Measurement Category   

The safety rating of the multimeter:   

1000 V, CAT I 

IEC Measurement Category I. The maximum measurable voltage is 1000 Vpk in the HI -LO 

terminal. 

600 V, CAT II   

IEC Measurement Category II. Inputs may be connected to AC mains power (up to 600 

VAC) under Category II overvoltage conditions.   

 

Measurement category definition   

 

Measurement CAT I

  applies to  measurements performed on circuits not directly 

connected to the AC mains. Examples are measurements on circuits not derived from the 

AC mains and specially protected (internal) mains- derived circuits. 

 

Measurement CAT II

  applies to protect against transients from  energy-consuming 

equipment supplied from the  fixed installation, such as TVs, PCs, portable tools,  and 

other household circuits. 

 

Measurement CAT III

   applies to protect against  transients in equipment in fixed 

equipment  installations, such as distribution panels,  feeders and short branch circuits, 

and lighting systems in large buildings.   

 

Measurement CAT IV

  applies to measurements performed at the source of the low- 

voltage installation. Examples are electricity meters and measurements on primary over 

current protection devices and ripple control units. 

 

 

 

Summary of Contents for 2203066

Page 1: ...VC 7055BT Digital Multimeter Quick Guide...

Page 2: ...erminals HI Sense and LO Sense Measurement Limits 4 Measurement Category 4 2 Quick Start 5 General Inspection 5 Dimensions 5 Foot Stool Adjustment 6 Front Panel Overview 6 Rear Panel Overview 8 User I...

Page 3: ...e instrument or other property might occur Safety Symbols Symbols on the Product The following symbol may appear on the product Direct current DC Warning risk of electric shock Alternating current AC...

Page 4: ...s product Refer to the user s manual for the Measurement Limits before connecting to the instrument Do not exceed any of the Measurement Limits defined in the following section Do not operate without...

Page 5: ...equency period capacitance diode and temperature test measurements Two Measurement Limits are defined for these terminals HI Input to LO Input Measurement Limit The Measurement Limit from HI Input to...

Page 6: ...ory II Inputs may be connected to AC mains power up to 600 VAC under Category II overvoltage conditions Measurement category definition Measurement CAT I applies to measurements performed on circuits...

Page 7: ...l You can check whether there is any loss of accessories with reference to this description If it is found that there is any accessory lost or damaged please get in touch with our distributor responsi...

Page 8: ...w 2 3 4 5 12 1 6 7 8 11 9 10 Figure 2 1 Front panel overview Item Name Description 1 LCD Display the user interface 2 Menu selection Keys Activate the corresponding menu 3 Operation Keys Save Collect...

Page 9: ...asurements 6 Range Direction Keys When the Range softkey is shown on the right menu you can press the key to switch between auto and manual range Press to enable manual range and increase or decrease...

Page 10: ...ng will be displayed in the Vice Display 12 Dual Exit Press the key to exit dual display mode Rear Panel Overview 1 3 2 5 4 Figure 2 2 Rear panel overview Item Name Description 1 RS232 Connect the PC...

Page 11: ...ay Power On 1 Connect the instrument to the 90 240 V AC supply using the supplied power cord Warning To avoid electric shock the instrument must be grounded properly 2 Press down the power button at t...

Page 12: ...trument damage do not discretionarily switch the measurement function when measuring DC Voltage Measurement DC Voltage AC Voltage Measurement AC Voltage DC Current Measurement DC Current AC Current Me...

Page 13: ...anel key to save current reading to the data table of the internal memory Auto record After setting number of points sample interval press the Start softkey to start recording After recording you can...

Page 14: ...ord 1 Configure the parameters Press the front panel key press the Auto softkey Press the Point softkey to specify the total number of readings to record The range is 1 to 1000 Press the Interval soft...

Page 15: ...l exists please contact us for our service 2 The reading does not change when a current signal is input 1 Check whether the test lead is correctly inserted into the current input terminals I terminal...

Page 16: ...prevent electrical shock disconnect the instrument from AC mains power and disconnect all test leads before cleaning 2 Clean the outside of the instrument using a wet soft cloth not dripping water Do...

Page 17: ...in personal injury or death Use only the correct fuse type Failure to do so could result in personal injury or instrument damage Voltage Fuse 90 240 V AC 250 V F1AL To perform the line fuse replacemen...

Reviews: