1.Safety Information
4
same voltage as the LO Input terminal, unless a current protection fuse is open.
Sense Terminals (HI Sense and LO Sense) Measurement Limits
The HI and LO sense terminals are used for four-wire resistance measurements.
The Measurement Limit from HI Sense to LO Input is 200 Vpk.
The Measurement Limit from HI Sense to LO Sense is 200 Vpk.
The Measurement Limit from LO Sense to LO Input is 2 Vpk.
Note
: The 200 Vpk limit on the sense terminals is the Measurement Limit. Operational
voltages in resistance measurements are much lower – up to ± 3 V in normal operation.
Measurement Category
The safety rating of the multimeter:
1000 V, CAT I
IEC Measurement Category I. The maximum measurable voltage is 1000 Vpk in the HI -LO
terminal.
600 V, CAT II
IEC Measurement Category II. Inputs may be connected to AC mains power (up to 600
VAC) under Category II overvoltage conditions.
Measurement category definition
Measurement CAT I
applies to measurements performed on circuits not directly
connected to the AC mains. Examples are measurements on circuits not derived from the
AC mains and specially protected (internal) mains- derived circuits.
Measurement CAT II
applies to protect against transients from energy-consuming
equipment supplied from the fixed installation, such as TVs, PCs, portable tools, and
other household circuits.
Measurement CAT III
applies to protect against transients in equipment in fixed
equipment installations, such as distribution panels, feeders and short branch circuits,
and lighting systems in large buildings.
Measurement CAT IV
applies to measurements performed at the source of the low-
voltage installation. Examples are electricity meters and measurements on primary over
current protection devices and ripple control units.