36
c) Fuse
•
A defective fuse must be replaced with a new fuse with the same specifications. Do not repair or bridge
a defective fuse, as this may cause a fire or result in fatal electric shock!
d) (Rechargeable) batteries
• Correct polarity must be observed while inserting the (rechargeable) battery.
• The (rechargeable) batteries should be removed from the device if it is not used for a long period of time
to avoid damage through leaking. Leaking or damaged (rechargeable) batteries might cause acid burns
when in contact with skin, therefore use suitable protective gloves to handle corrupted (rechargeable)
batteries.
• (Rechargeable) batteries must be kept out of reach of children. Do not leave (rechargeable) batteries
lying around, as there is risk, that children or pets swallow them.
• All (rechargeable) batteries should be replaced at the same time. Mixing old and new (rechargeable)
batteries in the device can lead to (rechargeable) battery leakage and device damage.
•
(Rechargeable) batteries must not be dismantled, short-circuited or thrown into fire. Never recharge
non-rechargeable batteries. There is a risk of explosion!
e) Multimeter
• Do not switch the meter on after it has been taken from a cold to a warm environment. The condensation
that forms might destroy the device. Allow the device to reach room temperature before switching it on.
• Check that the multimeter is set to the correct function before each measurement.
• To avoid electric shock, do not touch the connections/measuring points directly or indirectly during
measurement. Check the meter for damage before each measurement. Never carry out any
measurements if the protecting insulation is defective (torn, ripped off etc.)
• Do not use the during thunderstorms.
•
Never operate the product in direct proximity of strong magnetic or electromagnetic fields or transmitter
aerials or HF generators. This could affect measurement.
• For accurate measurements, replace the battery when the low battery symbol appears on the display.
• Pay attention to the information besides the warning signs at the test lead terminals. The measured
voltage or current must not exceed the values shown!
• Never operate the meter when the case is open. !DANGER TO LIFE!