28
11.5 Accumulated radiation dose rate
You can measure the accumulated radiation dose over time.
The initial unit is μSv. If the measured dose rate increases, the unit will automatically
change to mSv or Sv.
Press the
button to cycle through modes:
Infinite dose rate accumulation calculation mode, “μSv” will show.
Timing measurement mode (default 60 “060” mins), “TIME” will flash.
Stop timing measurement, “TIME” will stop flashing.
To set the measuring time:
1.
In timing measurement mode, press and hold
to set the measuring time
(001 - 999 minutes).
- Press ▲/ ▼to adjust values.
- Press
to toggle through settings.
2.
After all the values have been set, pressing
will save and exit settings.
11.6 Pulse counting
Mode
Description
CPS
cps
Pulse per second
cpm
Pulse per minute. Counter will automatically switch to Sv (1 Sv =
100 Rem).
Pulse
counting
Radiation is constantly measured by the Geiger tube and then
converted into a pulse rate.
“TIME” flashes when counting pulses, and stops flashing when
stopped counting pulses.
Press the
button to cycle through modes: “cps”→ “cpm” → “
”(counting)
→ “
” (stop counting).
Press
to exit pulse counting.