91
600 V overload protection
Fuses: µA/mA = 2x 0.55 A/240 V resettable, 1x F2 2.5 A/700 V ceramic
10 A = High-performance FF 10 A/600 V ceramic fuse
≤ 6 A = continuous measurement, >6 A = max. 30 secs at intervals of 15 minutes
The multimeter may display 3 counts when a measuring input is open
6
Alternating current (A/AC)
Range
Resolution
Accuracy
400.0 µA
0.1 µA
±(1.3% + 4)
4000 µA
1 µA
40.00 mA
0.01 mA
400.0 mA
0.1 mA
4.000 A
0.001 A
±(1.5% + 4)
10.00 A
0.01 A
±(1.8% + 7)
600 V overload protection
Fuses: µA/mA = 2x 0.55 A/240 V resettable, 1x F2 2.5 A/700 V ceramic
10 A = High-performance FF 10 A/600 V ceramic fuse
≤ 6 A = continuous measurement, >6A = max. 30 secs at intervals of 15 minutes
The multimeter may display 3 counts when a measuring input is open
TrueRMS peak (Crest Factor (CF)) ≤3 CF over the entire range
TrueRMS peak for non-sinusoidal signals plus tolerance
CF >1.0 - 2.0 + 3%
CF >2.0 - 2.5 + 5%
CF >2.5 - 3.0 + 7%