43
when measuring by basic accuracy that following conditions must be met:
• Ambient temperature: 23°C ± 5°C < 80%RH.
• Open and short calibration have been performed.
• D ≤ 0.1 for C, L measurements; Q ≤ 0.1 for R measurements.
• Do not measure when charge by AC adaptor. It may cause the reading rolling.
• See the operation manual for additional conditions.
Measurement tolerances
Statement of accuracy in ± (% of r display error in counts (= number of smallest points)).
The accuracy is valid for one year at a temperature of +23 °C (± 5 °C), and at a relative
humidity of less than 80%, non-condensing. Only valid if calibrated previously with open and
closed measurement input.
Dissipation factor „D“ ≤ 0.1 for C and L-measurements; Quality factor „Q“ ≤ 0.1 for R-
measurements.
Do not perform a measurement with the power adapter plugged in. This may cause the
displayed measurement values to fluctuate.
Inductivity
Range
100/120Hz
1kHz
10kHz
100kHz
Resolution
20.000 µH
N/A
N/A
N/A
0.5% + 30
[1]
0.001µH
200.00 µH
N/A
N/A
0.5% + 30
[1]
0.5% + 5
0.01µH
2000.0 µH
N/A
0.5% + 30
[1]
0.5% + 5
0.5% + 5
0.1µH
20.000 mH 0.5% + 30
[1]
0.5% + 5
0.5% + 5
1.0% + 5
0.001mH
200.00 mH 0.5% + 5
0.5% + 5
0.5% + 5
N/A
0.01mH
2000.0 mH 0.5% + 5
0.5% + 5
1.0% + 5
[2]
N/A
0.1mH
[1] The accuracy is indicated after subtraction of the offset inductiveness.
[2] <50 digit fluctuates
[3] If D > 0.1, the tolerance should be multiplied with
.