8 | Operation
LWT-100
18 / 26
B-H87.0.01.DB214-1.0
3. By briefly pressing the
Function key
, you can scroll through the parameters. Select
the parameter you would like to configure.
4. When you have selected the desired parameter, change the parameter to the de-
sired value with the
Up button
and the
Down button
.
5. The changes are saved after running through the entire
Configuration
menu.
STOR
appears in the display. The
Configuration
menu can be exited from any arbitrary
parameter by pressing and holding the
Function key
for 2 seconds. The changes
made up that point are saved.
Representation
Call up menu
Next parameter Change value
Save changes
Discard
changes
2s
Press: Single
step
Hold: Rapid
change
2s
2s
Outcome of an action
The
Configuration
menu is closed after the last parameter.
NOTE
If the product is switched off without saving the configuration, the last save value is re-
produced on the next start-up of the product.
8.2.3 Configuring parameters of the configuration menu
Description
The following representation shows the available parameters and various configura-
tion options.
Prerequisite
– The
Configuration
menu is open. See
Opening the configuration menu [
Instruction
1. Select the desired parameter you would like to configure.
2. Adjust the desired configuration in the selected parameter with the
Up button
and
Down button
.
3. The available configuration options are listed for each parameter in the following
representation.
Representation
Parameter
Values
Meaning
Input
INP
(OND
Measured variable - conductivity
SAL
Measured variable - salt content / salinity
TDS
Measured variable - total dissolved solids
Factor for TDS
CTDS
0.40 .. 1.00
Conversion factor for TDS measurement