background image

        

Legal notice

This is a publication by Conrad Electronic SE, Klaus-Conrad-Str. 1, D-92240 Hirschau (www.conrad.com). 
All rights including translation reserved. Reproduction by any method, e.g. photocopy, microfilming, or the capture in electronic data 
processing systems require the prior written approval by the editor. Reprinting, also in part, is prohibited. This publication represent the 
technical status at the time of printing.
© Copyright 2016 by Conrad Electronic SE.

G

Zero point calibration 

Take some pH7 buffer solution in a small container. The glass electrode should be completely covered 
during immersion. 
Turn on the meter by pressing the power button (3). 
Wait a while until the display has stabilised and then dip the electrode head in the buffer solution. Stir 
slowly with the meter in the buffer solution until the reading has stabilised. Leave the meter immersed and 
set the reading to a pH 7.00 using a calibration control “ZERO” (12). To do this, use the small alignment 
screwdriver.

   The pH changes slightly at a temperature deviation of 25 °C. The corresponding pH value is 

specified on the buffer solution container for the different temperatures. Use the corresponding 
value for alignment. 

Rinse the electrode with distilled water, rinse and pat the replaceable electrode dry. 

Reference point calibration 

The reference point shall always be close to the measured pH of the liquid. Select the reference point pH4 
if you want to measure acidic liquids (< pH7), and the reference point pH10 for alkaline liquids (> pH7).  
The reference buffer solution pH4 is supplied with the meter. 
Take some pH4 or pH10 buffer solution in a small container. The glass electrode should be completely 
covered during immersion.
Wait a while until the display has stabilised and then dip the electrode head in the buffer solution. Stir 
slowly with the meter in the buffer solution until the reading has stabilised. Leave the meter immersed and 
set the reading to a pH 4.00 (or pH 10.00) using the calibration control “SPAN” (10). To do this, use the 
small alignment screwdriver.

   The pH changes slightly at a temperature deviation of 25 °C. The corresponding pH value is 

specified on the buffer solution container for the different temperatures. Use the corresponding 
value for alignment.

Rinse the electrode with distilled water, rinse and pat the replaceable electrode dry.
Place the alignment screwdriver in the holder (11) within the battery compartment and close the battery 
compartment carefully.

MEASURING

Perform a calibration if the meter is not used for a long time or a electrode was changed. 
Proceed as follows for measuring:
Pull the protective cap (8) off the device.  This cap is merely plugged.
Screw the moistening cap (7) of the replaceable electrode (5) and check the liquid. The electrode must be 
moist. 
Clean the electrode with distilled water and tap it dry. Never touch the glass electrode (6) while doing so as 
this may damage it. 
Turn off the device by pressing the power button (3).
Dip the electrode into the liquid to be measured.  The glass electrode has to be completely covered by the 
liquid. 
Stir slowly with the meter in the buffer solution until a stable reading is displayed. Read this from the 
display. 
Turn off the device after the measurement by pressing the power button (3).
Clean the electrode and screw the moistening cap onto the device. If necessary, add a little soaking solu-
tion into the moistening cap.
Replace the protective cap (8) on the meter and stow it away carefully in the plastic box. 

   Measurements in hot liquids cause the electrode to wear much faster.

CHANGING THE ELECTRODE 

   The electrode is a wear part. Worn electrodes are not covered by the warranty/guarantee. 

If measurement is not possible or calibration cannot be carried out, the electrode is usually worn and must 
be replaced. 
To replace, proceed as follows: 
Make sure that the meter is switched off and the battery compartment is closed. 
Remove the protective cap (8) from the device. 
Loosen the screw fitting (4) of the electrode and take it off. 
Carefully pull the replaceable electrode (5) off the meter without turning the 
electrode. 
Insert the new replaceable electrode into the meter. Please note the notch in 
the plug region.  This fits correctly on the meter only on the right side. 
Screw the replaceable electrode again with the screw fitting (4). 
Check the liquid in the moistening cap and refill the soaking solution, if neces-
sary. 
Turn on the meter and check the operational readiness. A reading must be 
displayed. If no reading is displayed, please check the plug connection. Turn 
the device off. 
Perform a calibration before the first measurement. 
Replace the protective cap (8) on the meter and stow it away carefully in the 
plastic box.

TROUBLESHOOTING

In purchasing the measuring instrument, you have acquired a product with state-of-the-art design that is 
operationally reliable. Nevertheless, problems or faults may occur. Which is why we would like to describe 
here how you can solve any problems:

  

Always follow the safety instructions!

Error

Possible cause

Remedy

The meter cannot be 
switched on

The batteries are flat.

Change the batteries.

Incorrect readings

The temperature of the liquid is 
out of the allowable range of 0 to 
+ 50 °C. 

Bring the liquid within the given 
temperature range. 

The device has not been calibrated.

Perform a calibration. 

Calibration is not 
possible

The electrode is dirty. 

Clean the electrode. 

The buffer solutions are used. 

Use new buffer solutions. 

The electrode has dried up. 

Recover the electrode. 

The electrode is used up. 

Replace the electrode. 

   Repairs other than described above are carried out exclusively by authorised specialist staff. If 

you have questions concerning the use of the meter, please contact our technical support team.

CLEANING AND MAINTENANCE

Maintenance is limited to changing of batteries and electrodes.
Use a clean, lint-free, antistatic and slightly damp cloth to clean the device. Do not use any abrasive or 
chemical agents or detergents containing solvents.
The electrode must be cleaned after each measurement. Distilled water is sufficient for normal cleaning. 
A pepsin-cleaning fluid is required for measurements in protein-containing liquids (e.g. milk). Leave the 
electrode in the pepsin-cleaning fluid for about 5 minutes and then rinse thoroughly with distilled water. 
Pat the electrode dry and screw the moistening cap onto the electrode.

   All screw-fitted parts are protected from infiltration of water using rubber seals. Make sure that 

the seals are not pinched or damaged. Regular use of silicone grease is recommended.

Recovery of the glass electrode 

A dried glass electrode has a negative effect on the service life.  Dried electrodes may be recovered to 
render the open-pore glass layer permeable again. 
Dip the dried electrode in plenty of soaking solution (3 mol KCl solution) for min. 24 hours and then try 
re-calibrate. If this is not successful, extend the time again by 24 hours. Then, a measurement should be 
possible; if not, the electrode must be replaced.

DISPOSAL

   Electronic products are raw material and do not belong in the household waste. When the 

device has reached the end of its service life, please dispose of it, according to the current 
statutory requirements, at your local collecting site. It is prohibited to dispose of the device with 
domestic waste.

Disposal of used batteries!

 

 

As the end user, you are required by law (Battery Ordinance) to return all used batteries/re-
chargeable batteries; disposal of them in the household waste is prohibited!

 

 Batteries/rechargeable batteries containing harmful substances are marked with the following 
symbols, which indicate that disposal in the household waste is prohibited. The symbols of the 
critical heavy metals are: Cd = Cadmium, Hg = Mercury, Pb = Lead. You can return your used 
batteries/rechargeable batteries free of charge at the official collection points in your community, 
in our stores, or at places where batteries or rechargeable batteries are sold!

You thereby fulfil your statutory obligations and contribute to the protection of the environment!

TECHNICAL DATA

Display ...........................................................LCD
Measuring range ............................................0.01 - 14.00 pH
Accuracy ........................................................±0.01 + 1 count
Temperature compensation ...........................automatic, 0 to +50 °C
Power supply .................................................2 Lithium batteries (CR2032)
Weight ...........................................................approx. 85 g
Dimensions (H x Ø) in mm ............................168 x 37
Weight ...........................................................approx. 85 g

Summary of Contents for PHT-02 ATC

Page 1: ...echsel elektrode trocken Ber hren Sie nicht die Glaselektrode im inneren des Elektrodenkopfes da diese sonst besch digt wird D BEDIENUNGSANLEITUNG PH MESSSTICK PHT 02 ATC BEST NR 101137 BESTIMMUNGSGEM...

Page 2: ...auchen vollkommen bedeckt sein Schalten Sie das Messger t durch dr cken der Betriebstaste 3 ein Warten Sie etwas bis sich die Anzeige stabilisiert hat und tauchen Sie den Elektrodenkopf in die Puffer...

Page 3: ...able electrode dry Do not touch the glass electrode inside the electrode head as this may damage it G OPERATING INSTRUCTIONS PH PROBE PHT 02 ATC ITEM NO 101137 INTENDED USE The probe allows measuremen...

Page 4: ...lectrode 5 off the meter without turning the electrode Insert the new replaceable electrode into the meter Please note the notch in the plug region This fits correctly on the meter only on the right s...

Page 5: ...verre l int rieur de la t te de l lectrode car cela pourrait l endommager F MODE D EMPLOI PH M TRE PHT 02 ATC N DE COMMANDE 101137 UTILISATION CONFORME Le pH m tre permet de mesurer le pH de solutions...

Page 6: ...urner de l appareil de mesure Replacez la nouvelle lectrode de rechanger sur l appareil de mesure Notez l encoche dans la zone d insertion Elle permet d adapter l lectrode dans le bon sens sur l appar...

Page 7: ...t aan omdat deze anders beschadigd raakt O GEBRUIKSAANWIJZING PH MEETSTICK PHT 02 ATC BESTELNR 101137 BEOOGD GEBRUIK Met de meetstick kan de pH van waterige oplossingen binnen een bereik van 0 tot 14...

Page 8: ...ektrode 5 zonder deze van het meetinstru ment te draaien Plaats de nieuwe wisselelektrode op het meetinstrument Let op in het gedeelte waarmee u de elektrode plaatst zit een inkeping Deze past alleen...

Reviews: