47
Direct voltage
Range
Accuracy
Resolution
20.00 mV
±(1,2% + 8)
0.01 mV
200.0 mV
±(1,0% + 8)
0.1 mV
2.000 V
±(0.7% + 6)
0.001 V
20.00 V
0.01 V
200.0 V
0.1 V
600 V
1 V
Overload protection 600 V; Impedance: >10 MΩ (mV-range >1 GΩ)
Alternating voltage
Range
Accuracy
Resolution
20.00 mV
±(1,5% + 8)
0.01 mV
200.0 mV
0.1 mV
2.000 V
±(1,2% + 6)
0.001 V
20.00 V
0.01 V
200.0 V
0.1 V
600 V
±(1,5% + 4)
1 V
Specified measured range: 5-100% of the measured range
Permissible display with a short-circuited measuring input: <10 counts
Frequency range 45 – 400 Hz; overload protection 600 V
Measuring converter AC-V: AC-coupled, real effective value (True RMS), impedance: >10 MΩ, 100 pF
Additional tolerance at the following crest factor of the measuring signal:
CF 1.0 – 2.0
+3% for accuracy
CF 2.0 – 2.5
+5% for accuracy
CF 2.5 – 3.0
+7% for accuracy
Direct current
Range
Accuracy
Resolution
200.0 µA
±(1,0% + 6)
0.1 µA
2000 µA
1 µA
20.00 mA
0.01 mA
200.0 mA
0.1 mA
2.000 A
±(1,2% + 8)
0.001 A
10.00 A
0.01 A
Overload protection 600 V; high-performance ceramic fuses:
µA/mA-range F1: FF0, 2 A H 600 V (6 x 32 mm)
A-range F2: F10A H 600 V (6 x 25 mm)