Direct voltage (V/DC)
Range
Resolution
Accuracy
60.00 mV*
0.01 mV
±(1.2% + 8)
600.0 mV*
0.1 mV
±(1.0% + 8)
6.000 V
0.001 V
±(0.9% + 4)
60.00 V
0.01 V
600.0 V
0.1 V
*Only available in “mV” mode
Specified measurement range: 5 - 100% of the measurement range
600 V overload protection; Impedance: 10 MΩ (mV: ≤1000 MΩ)
The multimeter may display ≤10 counts if a measurement input is short-circuited.
Direct voltage (V/DC) LoZ
Range
Resolution
Accuracy
6.000 V
0.001 V
±(1.8% + 7)
60.00 V
0.01 V
600.0 V*
0.1 V
Specified measurement range: 5 - 100% of the measurement range
600 V overload protection; Impedance: 400 kΩ (*max. 250 V, 3s)
The multimeter may display ≤10 counts if a measurement input is short-circuited.
After using the LoZ feature, leave the multimeter for 1 minute before using it again
Alternating voltage (V/AC)
Range
Resolution
Accuracy
60.00 mV*
0.01 mV
±(1.5% + 4)
600.0 mV*
0.1 mV
6.000 V
0.001 V
±(1.3% + 4)
60.00 V
0.01 V
600.0 V
0.1 V
*Only available in “mV” mode
Specified measurement range: 5 - 100% of the measurement range
Frequency range: 45 - 400 Hz; Overload protection: 600 V; Impedance: 10 MΩ (mV: ≤1000 MΩ)
The multimeter may display 10 counts if a measurement input is short-circuited
TrueRMS peak (Crest Factor (CF)) ≤3 CF to 600 V
TrueRMS peak for non-sinusoidal signals plus tolerance:
CF >1.0 - 2.0 + 3%
CF >2.0 - 2.5 + 5%
CF >2.5 - 3.0 + 7%