background image

LED STATUS

Logger status

Battery

Signal LED

Power LED

switched on

---

flashes once when 

switching on

flashes once when 

switching on

switched off

---

off

off

records data**

normal

off

flashes*

low

flashes twice*

repeating time synchronisation

---

shines

flashes every ten 

seconds

records data without connection to 

network

normal

flashes twice*

flashes*

low

flashes twice*

r transmits data

normal

flashes (only data 

transmission)

flashes*

low

flashes twice*

r memory is full

normal

flashes three times*

flashes*

low

flashes twice*

data r transmits

---

flashes continuously

flashes*

not paired

normal

shines

flashes every ten 

seconds

low

flashes twice every ten 

seconds

pairing

normal

flashes six times within 

one second

flashes every ten 

seconds

low

flashes twice every ten 

seconds

resetting to factory settings / resetting 

network

---

off

flashes continuously for 

five seconds

low battery (product cannot be 

switched on)

low

flashes continuously

flashes continuously

*If the sampling rate is longer than one minute, the LED will flash every minute. Otherwise, the LED will flash 

during each sampling.
**If the sampling rate is two minutes or longer, the red power LED flashes once every minute (provided that 

battery capacity is normal and the product is connected to the network). During the minute the sampling 

takes place, the green signal LED will additional flash once.

When the logger has been paired with the dongle and is ready to use, and you press the button 

“CONNECT”, the red power LED will flash once. This serves to check the operational readiness of 

the logger.

TROUBLESHOOTING

Logger status

measure

repeating time synchronisation

Make  sure  the  dongle  is  properly  connected  to 

• 

the computer.
Make sure the software is running.

• 

Make sure that a network has been set up.

• 

If  required,  close  the  menu  item  “Sensor 

• 

Administration”.
Pair the product to the network. If this does not 

• 

work, reduce the distance to the dongle and try 

again.

records data without connection to network

r memory is full

battery low

Insert a new battery.

DISPOSAL

Product

a) 

Electronic devices are recyclable waste and must not be disposed of in the household waste!
At  the  end  of  its  service  life,  dispose  of  the  product  according  to  the  relevant  statutory 

regulations.
Remove any inserted (rechargeable) batteries and dispose of them separately from the product.

(Rechargeable) batteries

b) 
You  as  the  end  user  are  required  by  law  (Battery  Ordinance)  to  return  all  used  batteries/rechargeable 

batteries. Disposing of them in the household waste is prohibited!

Contaminated  (rechargeable)  batteries  are  labelled  with  this  symbol  to  indicate  that  disposal 

in the domestic waste is forbidden. The designations for the heavy metals involved are: Cd = 

Cadmium, Hg = Mercury, Pb = Lead (name on (rechargeable) batteries, e.g. below the trash 

icon on the left).
Used (rechargeable) batteries can be returned to collection points in your municipality, our stores 

or wherever (rechargeable) batteries are sold. 

You thus fulfil your statutory obligations and contribute to the protection of the environment.

 Legal notice

These operating instructions are a publication by Voltcraft®, Lindenweg 15, D-92242 Hirschau/Germany, Phone +49 180/586 582 7  

(www.voltcraft.de).
All rights including translation reserved. Reproduction by any method, e.g. photocopy, microfilming, or the capture in electronic data processing 

systems require the prior written approval by the editor. Reprinting, also in part, is prohibited.
These operating instructions represent the technical status at the time of printing. Changes in technology and equipment reserved.

© copyright 2011 by Voltcraft®

V3_0811_02-mk

TEcHNIcAL DATA

General
Operating voltage: 

3.6 V battery type ½ AA

Range: 

30 m (line of sight)

Transmission rate:

250 kbps

Internal memory:

512 kbits

Operating temperature:

-40 to +70 ºC

Operating humidity:

0 – 100 %

Storage temperature:

-10 to +50 ºC

Storage humidity:

10 – 85 %

Dimensions (W x H x D): 

28 x 24 x 141 mm

WDL-TH Temp/rH
Measuring range (temperature):

-40 to +70 ºC

Measuring range (humidity):

0 – 100 %

Accuracy (temperature):

±1.2 ºC (0 to +40 ºC)
±2.4 ºC (-40 to 0 ºC / +40 to +70 ºC)

Accuracy (humidity):

±3 % (20 – 60 %)
±4 % (10 – 20 % / 60 – 90 %)
±5 % (0 – 10 % / 90 – 100 %)

Resolution:

0.1 % / 0.1 ºC

Weight:

approx. 33 g

WDL-P Pressure
Measuring range:

300 – 1100 mbar

Accuracy:

±2.5 mbar (0 to +65 ºC / 700 – 1100 mbar)
±3 mbar (0 to +65 ºC / 300 – 700 mbar)
±4 mbar (-20 to 0 ºC / 300 – 1100 mbar)

Resolution:

0.01 mbar

Weight:

approx. 33 g

WDL-k Temp Type k
Measuring range:

-40 to +520 ºC

Accuracy:

±(0.5 % rdg + 2.9 ºC) at +25 ºC (typical)

Resolution:

0.1 ºC

Weight:

approx. 34 g

Summary of Contents for WDL-TH

Page 1: ...einschalten werden keine Daten aufgezeichnet bis das Produkt mit dem Netzwerk verbunden ist Das Produkt unternimmt automatisch alle zehn Sekunden einen Verbindungsversuch falls es bereits vor dem Bat...

Page 2: ...es nicht funktioniert verringern Sie die Distanz zum Dongle und versuchen Sie es erneut zeichnet Daten ohne Verbindung zum Netzwerk auf zeichnet auf Speicher ist voll Batterie schwach Legen Sie eine n...

Page 3: ...ork prior to the battery replacement Alternatively press the button CONNECT to establish a connection OPERATION The product only works in conjunction with the optionally available WDL PC Wireless Data...

Page 4: ...d waste At the end of its service life dispose of the product according to the relevant statutory regulations Remove any inserted rechargeable batteries and dispose of them separately from the product...

Page 5: ...des piles 3 Apr s avoir allum le produit aucune donn e ne sera enregistr e avant que le produit ne soit connect au r seau Si le produit tait reli au r seau avant le remplacement de la batterie il essa...

Page 6: ...e la cl lectronique et essayer nouveau enregistre des donn es sans connexion au r seau enregistre la m moire est pleine pile faible Ins rer une nouvelle pile elimination des dechets Produit a Les appa...

Page 7: ...het product is ingeschakeld worden er geen gegevens opgeslagen todat het product aan het netwerk is gekoppeld Het product voert automatisch om de tien seconden een verbindingspoging uit in het geval d...

Page 8: ...nder verbinding met het netwerk Slaat op geheugen is vol Batterij bijna leeg Plaats een nieuwe batterij verwijdering Product a Elektronische apparaten zijn recyclebare stoffen en horen niet bij het hu...

Reviews: