Power Analyzer User Manual
© Voltech 2010
Page 35
Ethernet communications are not available on the Basic Comms version of the
.
For basic communication needs via TCP/IP the user can try the Agilent
Connection Expert contained in the
Agilent IO Libraries Suite 15.0
The Ethernet mode (Static/DHCP), IP address, default gateway and subnet mask
are unchan
ged after a “*RST” or “:DVC” command.
4.9.
S
YSTEM
C
ONFIGURATION
.
Set up harmonics, distortion, the clock and auto zero.
To select a menu item, use the
keys and then press
for detailed options.
Harmonics Setup
For both voltage and current harmonics, a number of different parameters can
be set. These setting are independent of the mode that is currently selected.
Sequence: All or odd harmonics only
Range: The maximum harmonic (up to 50)
Format: Display harmonics as absolute values or as a percentage of the
fundamental (1st) harmonic.
Distortion Setup
For both voltage and current harmonics, a number of different parameters can
be set. These setting are independent of the mode that is currently selected.
For following settings can be made for distortion.
Formula: Series or difference (default = series formula)
Sequence: Include all harmonics or only odd harmonics in the series
formula. (default = all harmonics)
Range: The maximum harmonic to be included in the series formula.
(default = 7)
DC (H0): Include or exclude DC in the series formula. (default =
exclude)
Reference: rms or 1st harmonic. (default = rms)
See section 6.4 for details of the actual equations used.