R
EMOTE
O
PERATION
Page 42
© Voltech 2010
Power Analyzer User Manual
":HMX:THD:HZ <value>"
Choose to include or exclude THD
Harmonic zero:
<value> = 0
→ exclude
<value> = 1
→ include
":HMX:THD:DC <value>"
Select the THD reference:
<value> = 0
→ fundamental
<value> = 1
→ rms
“:INP:FILT:LPAS <value>”
Sets the low pass frequency filter state:
<value> = 0 -> Low Pass Filter Disabled
<value> = 1 -> Low Pass Filter Enabled
":INP:FILT:LPAS?"
Returns the state of the low pass filter.
":INT:START <value>"
Selects either Manual Start Method or
Clock Start Method.
<value> = 0 → Manual Start Method
<value> = 1 → Clock Start Method
":INT:MAN:RUN"
Starts integration when in Manual Start
Method. Requires integration mode
active, manual start selected and
integration not running.
":INT:MAN:STOP"
Stops integration when in Manual Start
Method. Requires integration mode active,
manual start selected and integration
running.
":INT:RESET"
Resets integration values. Requires
integration mode active and integration
not running.
":INT:CLK:TIME xx-xx-xxX"
Sets the start time for the integrator when
configured for Clock Start Method. Start
time sent in current time format.
xx-xx-xxX stands for hh-mm-ss
(uppercase „X‟ is not used) for 24hr time
format or hh-mm-ss(A or P) for AM/PM
time format.