IEC61000-3 S
OFTWARE FOR
W
INDOWS
U
SER
M
ANUAL
G
ETTING
S
TARTED
Voltech Instruments
VPN 98-025/7
Page 59
AND
- has EITHER a delayed restart (> 10's of seconds)
OR
- Manual restart after power supply interruption
(Choose 'Auto >2 / day or Manual Switching' to automatically set up
this limit.)
c) 7% for equipment that is:
- Attended whilst in use
OR
- Switched on automatically or manually no more than twice per day
AND
- has EITHER a delayed restart (> 10's of seconds)
OR
- Manual restart after power supply interruption
(Choose 'Attended or switching <= 2 /day' to automatically set up this
limit.)
The Pst limit is 1
The Plt limit is 0.65
The N value must be 12
The test duration must be 10 minutes for a Pst test, and should be 120
minutes (2 hours) for a Plt test.
(See the software help system for details of flicker test limits.)
When you have entered all the data, click on the OK button to return
to the Configure display shown above, and from that, click on the OK
button again to return to the Top Level display.