Applying the 2001 limits.
The various class limits are compared to the measurements obtained in the
following way:
(from 6.2.3.3)
1. The average value for the individual harmonic currents, taken over the
entire test observation period shall be less than or equal to the applicable
limits.
2. For each harmonic order, all 1.5s smoothed rms harmonic current values,
shall be less than or equal to 150% of the applicable limits.
3. Harmonic currents less than 0.6% of the input current measured under
the test conditions, or less than 5mA, whichever is greater, are disregarded.
4. For the 21st and higher odd order harmonics, the average values
obtained for each individual odd harmonic over the full observation period,
calculated from the 1.5s smoothed rms values may exceed the applicable
limits by 50% provided that:
•
the measured POHC does not exceed the POHC which can be
calculated from the applicable limits;
•
all 1.5 s smoothed rms individual harmonic current values shall be
less than or equal to 150% of the applicable limits.
In other words the equipment under test meets the limits if you:
1. Measure the harmonics 2 to 40 of the input current for each and every
consecutive 16-cycle (320mS @50Hz) block of the measurement and
smooth them with a 1.5 s time constant filter. For all these individual
values of data, disregard small values and compare to 150% of the relevant
Summary of Contents for PM3000ACE