If a luminaire has a built-in dimming device, the harmonic currents shall be
measured with the maximum load of the lamps as specified by the
manufacturer. The setting of the dimming device is varied in 5 equidistant
steps between the minimum and the maximum power in order to obtain
comprehensive results.
Ballasts and step-down converters
Ballast for fluorescent or other discharge lamps or step-down converters for
tungsten halogen or other filament lamps shall be tested with reference
lamps or with lamps having electrical characteristics close to their nominal
values. In case of doubt, measurements are made with reference lamps.
In the case where a ballast can be used with or without a series capacitor or
where a ballast or step-down converter is designed for several types of
lamps, the manufacturer shall indicate in his catalogue for which type of
circuit and lamps the ballast fulfils the harmonic requirements and the
ballast shall be tested accordingly.
Test conditions for independent and built-in incandescent lamp
dimmers
The independent dimmer is tested with incandescent lamps having the
maximum power allowed for the dimmer. The control is set to a firing
angle of 90° ± 5°, or if controlled by steps, to that step closest to 90°.
Test conditions for vacuum cleaners
The vacuum cleaner is tested with the air inlet adjusted according to the
normal operation as defined in 2.2.9 of IEC 60335-2-2. The control is set to
a firing angle of 90° ± 5°, or if controlled by steps, to that step closest to
90°.
Test conditions for washing machines