Series S34 Instruction Manual
Chapter 3 Operation
3-17
M-000-00030
Pipe O.D. in.
and
Pipe Wall in.
, are the dimensions of the pipe in which the transducers will be
installed and are required to calculate the flow rate. The Pipe I.D. does not need to be entered as
it is calculated by the handheld unit. Appendix C contains typical pipe wall and outside
diameter (OD) information.
Pipe Material
and
Liner Material
, if applicable, are the materials of the pipe in which the
transducers will be installed. Appendix C contains speed of sound information for various pipe
materials.
Transducer Type,
defines the frequency of the transducer being used for measurement of the
liquid inside the pipe. The three transducer frequencies to choose from include 0.5 MHz, 1
MHz, or 2 MHz.
TransTraverses,
defines the path of the ultrasonic sound wave that is being used for
measurement of the liquid inside the pipe. The selections include 1 path (Z), 2 path (V), or 4
path (W).
Spacing Temp
, is used in calculations to determine the distance (spacing) between the scribe
lines of the transducers. Enter the temperature of the liquid being measured at the time the
transducers are being installed for measurement.
Fluid Temp 1,
is the temperature of the liquid inside the pipe. The temperature can be entered
manually by selecting
Don’t Use RTD
. It can also be measured with a resistance temperature
detector (RTD) by selecting
Use RTD
. If
Use RTD
is selected and the meter display indicates a
temperature fault, a substitute value can be entered to allow flow calculations to continue at a
fixed value until the source of the fault is identified and corrected. This is accomplished by
selecting
Don
’t Use RTD
and entering the temperature of the liquid being measured. The units
of measure of the displayed values are the same as the units configured for the flow meter.
Note: if a V only model was purchased, select
Don’t Use RTD
and enter the temperature of the
fluid being measured.
Sound Speed SRC,
provides the user with three options for calculating the speed of sound of the
liquid being measured. Select
Spacing Temp
to use the temperature programmed for
determining the distance (spacing) between the scribe lines of the transducers to calculate the
speed of sound. Select
TOF_Calc
to use the measured time of flight (TOF) of the liquid being
measured to calculate the speed of sound. Select
Fluid Temp 1
to use either the programmed or
measured temperature of the liquid to calculate the speed of sound.
Summary of Contents for SonoPro S34 Series
Page 34: ...Series S34 Instruction Manual Chapter 3 Operation 3 5 M 000 00030 Output Menu...
Page 69: ...Series S34 Instruction Manual Appendix A Specifications A 2 M 000 00030...
Page 73: ...Series S34 Instruction Manual Appendix C Sound Speed and Pipe Data C 2 M 000 00030 Pipe Chart...
Page 74: ...Series S34 Instruction Manual Appendix C Sound Speed and Pipe Data C 3 M 000 00030...
Page 75: ...Series S34 Instruction Manual Appendix C Sound Speed and Pipe Data C 4 M 000 00030...
Page 76: ...Series S34 Instruction Manual Appendix C Sound Speed and Pipe Data C 5 M 000 00030...
Page 77: ...Series S34 Instruction Manual Appendix C Sound Speed and Pipe Data C 6 M 000 00030...
Page 78: ...Series S34 Instruction Manual Appendix C Sound Speed and Pipe Data C 7 M 000 00030...
Page 79: ...Series S34 Instruction Manual Appendix C Sound Speed and Pipe Data C 8 M 000 00030...