Series S34 Instruction Manual
Chapter 6 Troubleshooting and Repair
M-000-00030
6-6
•
Set Full Charge?
= YES or NO. If YES is selected,
then the “Battery State” will
change from the current value to a value of 100%. The default value is NO. As long
as this is selected the “Battery State” will not be reset.
•
SR Coefficient
= Flow rate multiplier. This parameter has a default value of 1. It
should not be changed unless directed by the factory.
•
Polarity
= Polarity of the drive signal (i.e. charge on the piezoelectric crystal).
•
Num Cycles
= The number of sound wave pulses that are being sent through the pipe
and fluid and received by the transducers.
•
Waveform
= Unkeyed or keyed.
•
Manual Flight uS
= A manual adjustment for the Time of Flight. Zero will use the
value calculated by the application (i.e. value calculated from transducer, pipe size,
fluid, setup temp, etc.) This is used when the received waveform is not in the
sampling window, especially the wave front.
•
Fire
= Factory use only.
•
Reading Flow?
= YES or NO.
•
Num.
Zero Loops =
Number of samples (loops) of up/down waveforms, taken at no
flow, to calculate the
Zero Cal.
average found in the
Calibration
menu.
•
RESET Bluetooth =
YES or NO. Default value is YES.
•
Manual Gain/Pwr =
YES or NO. Default value is NO.
•
Wave Simulation =
This feature simulates actual waveforms but is for demonstration
purposes only. If there are simulation files stored in the flash, it will have a selection
other than
No Simulation
. Normal operation resumes if the power is cycled or
No
Simulation
is selected. Units operating as a flow meter should never use this feature.
Simulation reprograms the application. Simulation files should be provided for
training and sales purposes only.
Summary of Contents for SonoPro S34 Series
Page 34: ...Series S34 Instruction Manual Chapter 3 Operation 3 5 M 000 00030 Output Menu...
Page 69: ...Series S34 Instruction Manual Appendix A Specifications A 2 M 000 00030...
Page 73: ...Series S34 Instruction Manual Appendix C Sound Speed and Pipe Data C 2 M 000 00030 Pipe Chart...
Page 74: ...Series S34 Instruction Manual Appendix C Sound Speed and Pipe Data C 3 M 000 00030...
Page 75: ...Series S34 Instruction Manual Appendix C Sound Speed and Pipe Data C 4 M 000 00030...
Page 76: ...Series S34 Instruction Manual Appendix C Sound Speed and Pipe Data C 5 M 000 00030...
Page 77: ...Series S34 Instruction Manual Appendix C Sound Speed and Pipe Data C 6 M 000 00030...
Page 78: ...Series S34 Instruction Manual Appendix C Sound Speed and Pipe Data C 7 M 000 00030...
Page 79: ...Series S34 Instruction Manual Appendix C Sound Speed and Pipe Data C 8 M 000 00030...