XR3D-6 Maintenance & Troubleshooting
EM-XR3D-6-001-EN.D
3-13
©2011 VOTI Inc.
3.5 Seasoning the X-ray Generator Tube
If the XR3D has been powered down for a period longer than 3 days, it is necessary to season
the X-ray generator tube to ensure it is still good. To season the X-ray generator tube, proceed
as follows:
1. A VOTI-certified technician logs on using Administrator permission to access the Windows
TM
desktop.
2. Click on DIOX shortcut (
c:\Program Files\VOTI\BioSansXDAQ\DIOX.exe
) to start
the DIOX application.
The DIOX window appears.
3. Press the XRAY button (
).
The lights on the top of the scanner will turn red. The DIOX indicators for Red/Green Lights
and X-Ray On/Off will also turn red.
It means that the source is enabled to generate X-Rays but it is not powered. Since the
source is not powered, there is no radiation emitted.
4. Start VOTI's Diagnostic Tool application, which is the software for the X-ray source, located
at
C:\MonoBlock XP\X3825 Spellman*.exe
.