XR3D-6 Maintenance & Troubleshooting
3-16
EM-XR3D-6-001-EN.D
©2011 VOTI Inc.
The DIOX window appears.
3. Press the XRAY button (
).
The lights on the top of the scanner will turn red. The DIOX indicators for Red/Green Lights
and X-ray On/Off will also turn red.
It means that the source is enabled to generate X-rays but it is not powered. Since the
source is not powered, there is no radiation emitted.
4. Start VOTI's Diagnostic Tool application, which is the software for the X-ray source, located
at
c:\MonoBlock XP\X3825 Spellman*.exe
.
When doing testing, never start at the highest kV and the highest Current.
5. In VOTI's Diagnostic Tool application:
a. Click
Program Voltage
to set the anode voltage.
The unit is 0.1 KV. The first test should be run at 100 KV, enter
1000
and click
OK
.