XR3D-6 Maintenance & Troubleshooting
EM-XR3D-6-001-EN.D
3-17
©2011 VOTI Inc.
b. Click
Program Current
to set the tube current.
The unit is 1 µA. The first test should be run at 200 µA, enter
200
and click
OK
.
6. Click
XrayOn
to generate X-Rays.
The
HV Monitor
turns from green “HV OFF” to red “HV ON”.
If the HV Monitor box does not turn red:
o
Click
Enable WDT
to toggle the display to
Disable WDT
.
o
Click
Disable WDT
to toggle the display back to
Enable WDT
.
o
Click
XrayOn
.
3.6.1 Measuring the X-rays
Using a Dosimeter, position the sensor 5cm from the XR3D-6 and measure all external surfaces,
beginning with the front Access Panel.
1. Use a certified and calibrated survey dosimeter.
2. Write the model and serial number of the device on the Test Record Sheet.
3. Take measurement from 5 cm of any external surface* of the scanner
Summary of Contents for XR3D-6
Page 2: ...XR3D 6 Maintenance Troubleshooting ii EM XR3D 6 001 EN D 2011 VOTI Inc...
Page 6: ......
Page 20: ...XR3D 6 Maintenance Troubleshooting 2 8 EM XR3D 6 001 EN D 2011 VOTI Inc...
Page 32: ...XR3D 6 Maintenance Troubleshooting 3 12 EM XR3D 6 001 EN D 2011 VOTI Inc...
Page 48: ...XR3D 6 Maintenance Troubleshooting 3 28 EM XR3D 6 001 EN D 2011 VOTI Inc...
Page 52: ...XR3D 6 Maintenance Troubleshooting 4 4 EM XR3D 6 001 EN D 2011 VOTI Inc...