XR3D-6 Maintenance & Troubleshooting
EM-XR3D-6-001-EN.D
3-21
©2011 VOTI Inc.
10. Repeat this procedure for each step wedge.
11. After completing the test, remove the tape placed over the sensor inside the tunnel.
3.7.1 Re-Calibrate the XR3D
If there is a difference of more than 5%, redo the calibration and update the XData files in
c:\Program Files\VOTI\BioSansXDAQ\Input\BIOSANS\XData
. To use the Calibration
application, proceed as follows:
1. Start the Calibration application by, clicking
C:\Program Files\VOTI\BioSansXDAQ\
CalibrationX1-V2.exe
.
2. Click
Load First Plexiglass Archive
to load the first Plexiglass scan from the archive folder.
3. Enter the detector range in
Top
and
Bottom
to define the detectors that are being used, for
example,
Top
: 0
Bottom
: 251.
This value is based on the location of the item in the archived scan image.