XR3D-6 Maintenance & Troubleshooting
EM-XR3D-6-001-EN.D
3-27
©2011 VOTI Inc.
11. Try to get the best diode plot if it does not look like this image.
This is done by accessing the X-Ray source by removing the side access panel. Loosen the
bolts securing the X-ray Source base and move the source back or forward a millimeter or
two (
not much movement is required
). After adjusting the Source base, secure the bolts, and
re-do the test.
NEVER move the source while it is generating X-rays. Always stop the source to do the
adjustment.
ALWAYS adjust the source by loosening the bolts at the source base support.
ALWAYS wait 5 minutes if removing power from the source to allow the capacitor to
discharge before unplugging the source.