XR3D-6 Maintenance & Troubleshooting
6-4
EM-XR3D-6-001-EN.D
©2011 VOTI Inc.
10. Lift out the computer and package to send back to VOTI Support.
11. Insert a new computer and replace and secure the side bolts and the stiffener bars.
12. Re-attach the detector ribbon connectors and secure the computer top access panel using 2
screws.
13. Plug in the 15-pin connector and the connectors to the monitor and the UPS.
14. Reapply AC power and turn back on the UPS.
15. Turn on the computer using the left base toggle switch (if it does not turn on automatically).
16. Replace and secure the top access panel and the base front and back access panels.
The system restarts.
6.3 Replace the Detectors
To access the Detectors, the Rear Base Panel, Rear Tunnel Cover, Tunnel Top Access Panel,
the Detector Top or Side Cover must be removed.
Tools:
Screwdriver with Security Chuck
10mm Hex head socket wrench
Work Gloves
1. Unplug the scanner power cable from the wall receptacle.
2. Using a screwdriver with the security chuck, remove the (2 or 4) security screws (M6 x 25
SS machine screw with a 6 point security head) on the base front (to access the UPS) and
rear access panel.
3. Turn off the UPS by pressing the power button.
4. Using a screwdriver with the security chuck, remove the eight (8) security screws (M6 x 25
SS machine screw with a 6 point security head) on the two (2) top access panels.
Summary of Contents for XR3D-6
Page 2: ...XR3D 6 Maintenance Troubleshooting ii EM XR3D 6 001 EN D 2011 VOTI Inc...
Page 6: ......
Page 20: ...XR3D 6 Maintenance Troubleshooting 2 8 EM XR3D 6 001 EN D 2011 VOTI Inc...
Page 32: ...XR3D 6 Maintenance Troubleshooting 3 12 EM XR3D 6 001 EN D 2011 VOTI Inc...
Page 48: ...XR3D 6 Maintenance Troubleshooting 3 28 EM XR3D 6 001 EN D 2011 VOTI Inc...
Page 52: ...XR3D 6 Maintenance Troubleshooting 4 4 EM XR3D 6 001 EN D 2011 VOTI Inc...