VTI Instruments Corp.
58
EX2500A Web Page Operation
•
Module Options
: this field allows the user to enable or disable specified EX2500A features.
o
RAM ECC
: if enabled, the EX2500A will perform memory error checking and
correction (ECC). This option may slow down performance, but increases reliability.
o
Check Backplane Clocks on Startup
: if enabled, the EX2500A will check the 10 MHz
and 16 MHz backplane clocks on startup. If these clocks are not within their
specifications, an error code will be generated on the network LEDs and the emergency
backup OS will start instead of the full system.
o
PPS Output
: if enabled, the EX2500A will generate a single pulse-per-second (PPS)
waveform on its the external trigger output.
•
Change Password
: this user-configurable field can be used to change the password of the
EX2500A. Passwords are case-sensitive and the default password is
ex2500
. When a
configuration page is accessed after the password has been changed, a login prompt will
appear.
T
IME
The
Time
page allows the user to configure the EX2500As time settings. The time can be set
manually or may be synchronized with an IEEE 1588 source.
F
IGURE
5-7:
T
IME
P
AGE
•
Zone
:
allows the user to select an appropriate time zone.
•
Date
: allows the user to set the EX2500As date by entering the month (MM), day (DD), and
year (YYYY), respectively.
•
Time
: allows the user to set the EX2500As time by entering the hour (HH), minute (MM),
and second (SS), respectively. If the EX2500A is set as an IEEE 1588 Master, the hour should
be entered in using the 24-hour format.
•
PTP status
: indicates the IEEE 1588 clock status.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com