VTI Instruments Corp.
60
EX2500A Web Page Operation
•
LXI Module to Module Parameters
o
LXI Domain
: an 8-bit number, 0 to 255, which indicates the domain the EX2500A is
using for LAN events. All events sent by this device will include this data. Likewise, the
device will only accept events that include the same LXI Domain data.
•
Wired Trigger Parameters
o
Wired-OR Bias
: indicates whether the Wired-OR bias for each LXI trigger line (LXI0
through LXI7) is enabled or disabled.
LXI
LAN
E
VENTS
The
LXI LAN Events
page is used for module-to-module communication over Ethernet. The
LAN Events
menu is used to configure the EX2500A’s LAN Event input filters.
F
IGURE
5-9:
LAN
E
VENTS
C
ONFIGURATION
W
EB
P
AGE
•
Domain
: displays the current domain byte.
•
Log Enable
: checking this box allows the user to log LAN Events. The log can be viewed by
clicking on the
LAN Event Log
link in the
Command Menu
.
•
Clear Log
: checking this box will clear all events that are currently recorded in the LAN
Event Log.
•
Source
: indicates the LAN trigger source whose properties will be configured.
•
Event ID
(default settings shown): allows the user to specify an alternative LAN Event ID
name which is application-specific (e.g. “DoneWaiting”). This is defined in
IVI-3.15:
IviLxiSync Specification, 4.2.14 Trigger Source EventId
. Note that, if the Event ID is altered,
the default Event ID will no longer be accepted unless specifically returned to its default
name.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com