RANGe:DELay:AUTOmatic
Purpose
Enables the default delay times.
Type
Setting
Command Syntax
RANGe:DELay:AUTOmatic <boolean>
Command Parameters
<boolean> = 0 | 1 | OFF | ON
*RST Value
1
Query Syntax
RANGe:DELay:AUTOmatic?
Query Parameters
N/A
Query Response
0 | 1
Description
This command enables or disables the default delay times without having to perform a
unit reset. Default times are as follows:
Function
Range
Time
Voltage (dc)
ALL
5 ms
Voltage (ac)
ALL
AC Filter settling time
1
2- & 4-wire ohms
20
Ω to 200 kΩ
5 ms
2 MΩ
50 ms
20 MΩ
500 ms
Current (dc)
ALL
5 ms
Current (ac)
ALL
AC Filter settling time
1
Low
Mid
High
1
AC Filter settling time
800 ms
250 ms
60 ms