22
DNA/Protein Mode
1. Enter DNA/Protein Mode
Main menu, press
or
,
to select DNA/Protein Mode and press
to confirm.
2. Setup
parameters
Press
to set coefficients, input all the values of f1 to f4 using the numeric keypad according the
on-screen indications and press
to confirm.
3. Choose measure method
Press
to set method. Press
,
to choose Absorbance Difference 1 or Absorbance Difference 2,
then press
to confirm. If you do not want to measure reference, use
,
to select No, then press
to confirm.
4. Set concentration unit
Press
to set concentration unit. Use
,
to select unit and press
to confirm.
5. Measure
samples
Put the sample to be measured in the main light path and put reference in the reference light path, press
to measure. The result will be displayed in the data sheet.