Part 2: Software
Functions
This section introduces the functions of the UV-Vis Analyst.
Main Functions
Single wavelength photometric measurement
Go to a desired wavelength quickly and conveniently.
Photometric value display mode can be changed (%Transmittance or Absorbance).
Fixed points measurement
Multi-wavelength photometric measurement
Up to 20 wavelength points can be set.
Results will be grouped into a table format automatically.
Concentration measurement
2 methods to set the regression curve.
Up to 20 standards to set the regression curve. The UV-Vis Analyst will calculate the working curve
using a linear equation that fits the data. Enter factor values to generate regression curve.
3 methods for curve fit.
Linear fit, quadratic fit and cubic fit.
Wavelength scanning
Allows user to set scan step (0.1, 0.2, 0.5, 1.0 and 5.0 nm).
Spectrum display mode can be changed (Wavelength-%Transmittance or Wavelength-Absorbance).
Peaks and valleys will be automatically detected after scanning (user can define the peak threshold).
Powerful spectrum processing functions are provided.
Time scanning
Allows user to set scan interval (0.5, 1.0, 2.0, 5.0, 10, 30 and 60 s).
Spectrum display mode can be changed (Time-%Transmittance or Time-Absorbance).
Peaks and valleys will be automatically detected after scanning (user can define the peak threshold).
Powerful spectrum processing functions are provided.
DNA/Protein measurement
Wavelength points and ratios can be set up.
Results will be grouped into a table format automatically.