vwr.com I Instruction manual VWR® U Water System
dispense button to stop dispensing until the tank is refilled.
3.3.3 Degassing the System
a) Remove the stopper from the UP dispenser outlet.
b) Screw on the barbed adapter (in the accessory bag) to the dispenser outlet. Connect the PVC
tubing inside the accessory bag; insert one end to the adapter and put the other end into the
drain.
c) Press the dispense button to dispense water. Gas inside the system will be released out of the
system.
d) Wait till no more gas bubbles coming out of the tubing. Press the dispense button again to
stop degassing. Screw off the adapter.
3.3.4 Installing the Final Filter
a) Open the attached final filter package.
b) Screw the final filter onto the remote dispenser until finger-tight (no leak at dispensing). Do not
over tighten it as it may damage threads.
c) Press the dispense button to flush out gas in the filter.
d) Press the dispense button again to stop dispensing.
3.3.5 Tank Recirculation Setup (Optional)
For tank recirculation model only, swipe the screen to page 3 in Standby mode. Page 3 is
Maintenance Menu
. Tap
Setting
to enter setup menu. For system with T Pack, check
“
√
in the
red box before
Tank Recirculation.
Notes: Maintenance menu is password protected for system administrators. The log-in
information is in the USB stick shipped with the system.
This completes the system hardware installation.