60
vwr.com I Instruction manual VWR® U Water System
Warning!
Ultraviolet (UV) radiation is harmful to the eyes and skin. Do not observe the lamp
directly when it is illuminated. This system is equipped with a lamp cover to prevent
UV light leakage. This cover must be on ALL TIMEs when a UV lamp is installed.
Caution!
Keep the UV lamp straight in and out of the stainless steel chamber during its
installation to avoid any action that could cause the lamp to break.
a) Power off the VWR main system.
b) Screw off the nut on the back of the controller. Remove the controller and carefully remove the
used lamp from the chamber. Unplug the UV lamp base.
c) Wear gloves included in the new UV lamp package. Avoid direct skin contact with the quartz
glass of the UV lamp. Install the new UV lamp to the lamp base, and then gently insert the
new UV lamp completely into the chamber.
d) Screw the sealing cap on the tank cap, and then screw the nut to fix the controller on the tank
cap.
e) Turn on the power and operate the system. Scan the RFID tag for the UV sanitization module
on the RFID reader position ( ) to register it.
5.4.5 Replacing the Tank Vent Filter
a)
Remove the old tank vent filter.
b)
Screw the new tank vent filter to the bass on the tank.
c)
Scan the RFID chip for the tank vent filter on the system. Enter Components Installation menu
to reset the corresponding tank vent filter.
Nut
Sealing Cap
Tank Cap
Controller
Support