Use of replacement parts other than VWR original replacement parts
•
Electrical safety analysis not performed after repairs
2.6 Operating instructions by operator
Depending on the application and location of the chamber, we recommend that the operator of the freezer
provides the relevant information for safe operation of the chamber in a set of operating instructions.
Keep this set of operating instructions with the chamber at all times in a place where they
are clearly visible. They must be comprehensible and written in the language of the employ-
ees.
2.7 Measures to prevent accidents
The operator of the chamber must observe the locally applicable regulations for operating the chamber (for
Germany: Occupational Safety Regulations. Operation of refrigeration chambers, heat pumps and cooling
systems, GUV-R 500 chap. 2.35) and take precautions to prevent accidents.
Following measures have been taken by the manufacturer in order to prevent ignition and explosions:
•
Indications of the type plate
See operating manual chap. 1.6
•
Operating manual
An operating manual is available for each chamber.
•
Temperature monitoring
The chamber has a temperature display which can be read from outside.
An additional temperature safety device is built into the chamber. A visual and an audible signal (buzzer)
show exceeding of the temperature.
•
Safety, measurement and control devices
The safety, measuring, and control devices are easily accessible.
•
Electrostatic charge
The interior parts are grounded.
•
Non-ionizing radiation
Non-ionizing radiation is not intentionally produced, but released only for technical reasons by electrical
equipment (e.g. electric motors). The machine has strong permanent magnets. If persons with active
implants (e.g. pacemakers, defibrillators) keep a safe distance (distance of field source to implant) of 30
cm, an influence of these implants can be excluded with high probability.
•
Protection against touchable surfaces
Tested according to EN ISO 13732-3:2008.
•
Floors
See operating manual chap. 4.5 for installation
•
Cleaning
See operating manual chap. 22.
•
Examinations
UL chambers only: The chamber is certified by Underwriters Laboratories Inc.
®
according to the stand-
ards UL 61010-1, 3
rd
Edition, 2012-05, Rev. 2016-04; UL 61010-2-011, 1
st
Edition, 2017-01; CAN/CSA-
C22.2 No. 61010-1, 3
rd
Edition, 2012-05, Rev. 2016-04