vwr.com I Instruction manual VWR
®
ULT Freezer 352/528 Eco Premium 02/2022
Page 79/86
If recycling is not possible, all packing parts can also be disposed of with normal waste.
25.2 Equipment disposal
This equipment is marked with the crossed out wheeled bin symbol to indicate
that this equipment must not be disposed of with unsorted waste.
Instead it's your responsibility to correctly dispose of your equipment at lifecycle -
end by handling it over to an authorized facility for separate collection and recy-
cling. It's also your responsibility to decontaminate the equipment in case of bio-
logical, chemical and/or radiological contamination, so as to protect from health
hazards the persons involved in the disposal and recycling of the equipment.
For more information about where you can drop off your waste of equipment, please contact your local
dealer from whom you originally purchased this equipment.
By doing so, you will help to conserve natural and environmental resources and you will ensure that your
equipment is recycled in a manner that protects human health. Thank you.
The refrigerants used R290 (propane, GWP 3) and R170 (ethane, GWP 6) are inflammable
at ambient pressure. A suction is not required. Ensure the compliance with the applicable le-
gal requirements regarding qualification of staff, and documentation
26. Technical description
26.1 Factory calibration and adjustment
The chambers were calibrated and adjusted in factory. Calibration and adjustment were performed using
standardized test instructions, according to the QM DIN EN ISO 9001 system applied by the manufacturer.
All test equipment used is subject to the administration of measurement and test equipment that is also a
constituent part of the manufacturer’s QM DIN EN ISO 9001 system. They are controlled and calibrated to
a DKD-Standard at regular intervals.
Repeated calibrations are recommended in periods of 12 months. Please contact Avantor
Services for details.
26.2 Over current protection
The devices are equipped with an internal fuse not accessible from outside. If this fuse is blown, please
inform an electronic engineer or Avantor Services.
26.3 Technical data
Chamber size
(referring to the max. quantity of 2” boxes)
352
528
Exterior dimensions
Width, gross (including hinges, door handle, and controller
housing)
mm /
inch
920 /
36.2
1204 /
47.4
Height, gross (incl. castors)
mm /
inch
1966 /
77.4
1966 /
77.4
Depth, gross (including power connection, without door
handle and controller housing (equals depth when door
open))
mm /
inch
850 /
33.5
850 /
33.5
Depth, gross (including power connection, door handle
and controller housing)
mm /
inch
1005 /
39.6
1005 /
39.6
Wall clearance rear (minimum)
mm /
inch
100 /
3.9
100 /
3.9
Wall clearance side (side without hinges) (minimum)
mm /
inch
100 /
3.9
100 /
3.9
Wall clearance side (side with hinges) (minimum)
mm /
inch
245 /
9.6
245 /
9.6