The parts bag, located inside the cabinet, contains the following parts.
Install the bolts into the pre-tapped holes on the back of the freezer. Install
a neoprene cap on each bolt. Refer to Figures 1-3 and 1-4 for the locations
of the pre-tapped holes.
See Figures 1-3 and 1-4 for the location of the remote alarm contacts. The
remote alarm connector is located in the parts bag provided with the
manual. It must be installed if connecting the freezer to an alarm system.
After installing the wiring from the alarm system to the connector, install
the connector to the freezer microboard and secure with the two screws
provided. The remote alarm provides a NO (normally open) output, a NC
normally closed) output and COM (common). The contacts will trip on a
power outage, high temperature alarm or low temperature alarm. They will
also trip on high stage, control probe and microboard failures. Figure 1-10
shows the remote contacts in alarm state.
ULT Freezer 1-7
VWR International
Section 1
Installation and Start-up
Install Wall Bumpers
Quantity
Stock #
Description
Purpose
2
510016
1/4-20 x 5-1/2” Bolt
Wall Bumper
2
380520
Neoprene Cap
Cap Protector
Remote Alarm Contacts
IMPORTANT USER INFORMATION
IMPORTANT USER INFORMATION
Caution! Stored product should be protected
by a redundant 24 hour/day monitoring
system with alarm capability. An interconnect
jack and thermocouple are installed
for centralized monitoring,should on-board
system fail.
Figure 1-10.
Remote Alarm Contacts
Summary of Contents for 5700 Series
Page 19: ......
Page 21: ......
Page 41: ......
Page 44: ...ULT Freezer 7 1 VWR International Section 7 Parts ...
Page 45: ...7 2 ULT Freezer VWR International Section 7 Parts ...
Page 46: ...ULT Freezer 7 3 VWR International Section 7 Parts ...
Page 59: ......
Page 60: ...ULT Freezer 8 1 VWR International Section 8 Refrigeration Schematics ULT Freezer 8 1 ...
Page 61: ...8 2 ULT Freezer VWR International Section 8 Refrigeration Schematics ...
Page 62: ...ULT Freezer 8 3 VWR International Section 8 Refrigeration Schematics ...
Page 63: ...8 4 ULT Freezer VWR International Section 8 Refrigeration Schematics ...
Page 64: ...ULT Freezer 8 5 VWR International Section 8 Refrigeration Schematics ...
Page 65: ......
Page 66: ...ULT Freezer 9 1 VWR International Section 9 Electrical Schematics ...
Page 67: ...9 2 ULT Freezer VWR International Section 9 Electrical Schematics ...
Page 68: ...ULT Freezer 9 3 VWR International Section 9 Electrical Schematics ...
Page 70: ...ULT Freezer 9 5 VWR International Section 9 Electrical Schematics ...
Page 71: ...9 6 ULT Freezer VWR International Section 9 Electrical Schematics ...
Page 72: ...ULT Freezer 9 7 VWR International Section 9 Electrical Schematics ...
Page 73: ...9 8 ULT Freezer VWR International Section 9 Electrical Schematics ...
Page 74: ...ULT Freezer 9 9 VWR International Section 9 Electrical Schematics ...
Page 75: ...9 10 ULT Freezer VWR International Section 9 Electrical Schematics ...
Page 76: ...ULT Freezer 9 11 VWR International Section 9 Electrical Schematics ...
Page 77: ...9 12 ULT Freezer VWR International Section 9 Electrical Schematics ...
Page 78: ...ULT Freezer 9 13 VWR International Section 9 Electrical Schematics ...
Page 79: ...9 14 ULT Freezer VWR International Section 9 Electrical Schematics ...
Page 80: ...ULT Freezer 9 15 VWR International Section 9 Electrical Schematics ...
Page 81: ......
Page 90: ...VWR International P O Box 5015 Bristol CT 06011 8960 ...