3-2
ULT Freezer
VWR International
Section 3
Alarms
The "high stage system failure" condition is created when the high stage
compressor and fans run for 30 minutes and are not capable of cooling the
interstage heat exchanger to the proper temperature. Under this condition,
the high stage compressor and fans will turn off after 30 minutes and an
audible and visual alarm will occur. The audible alarm can be silenced and
will ring back every 15 minutes.
The microprocessor in Model 5700 series freezers continually scans all
probes including the control probe, heat exchanger probe and condenser
probe to ensure that they are operating properly. Should an error be
detected, the "Probe Failure" alarm will occur as described in Section 3.1
above. If an error is detected with the control probe, the high and low
stage compressors will run continuously. As a result, the cabinet
temperature will decrease until it reaches the lowest temperature that the
refrigeration system can maintain. If an error is detected with the heat
exchanger probe, the freezer will cycle properly at its temperature set point
using a 5 minute step start between the high and low stage compressors. If
an error is detected with the condenser probe, there is no impact on the
performance of the freezer; however, the hot condenser alarm may also
occur. Contact the VWR Service department or your local distributor.
Probe Failure Alarm
High Stage System
Failure
Summary of Contents for 5700 Series
Page 19: ......
Page 21: ......
Page 41: ......
Page 44: ...ULT Freezer 7 1 VWR International Section 7 Parts ...
Page 45: ...7 2 ULT Freezer VWR International Section 7 Parts ...
Page 46: ...ULT Freezer 7 3 VWR International Section 7 Parts ...
Page 59: ......
Page 60: ...ULT Freezer 8 1 VWR International Section 8 Refrigeration Schematics ULT Freezer 8 1 ...
Page 61: ...8 2 ULT Freezer VWR International Section 8 Refrigeration Schematics ...
Page 62: ...ULT Freezer 8 3 VWR International Section 8 Refrigeration Schematics ...
Page 63: ...8 4 ULT Freezer VWR International Section 8 Refrigeration Schematics ...
Page 64: ...ULT Freezer 8 5 VWR International Section 8 Refrigeration Schematics ...
Page 65: ......
Page 66: ...ULT Freezer 9 1 VWR International Section 9 Electrical Schematics ...
Page 67: ...9 2 ULT Freezer VWR International Section 9 Electrical Schematics ...
Page 68: ...ULT Freezer 9 3 VWR International Section 9 Electrical Schematics ...
Page 70: ...ULT Freezer 9 5 VWR International Section 9 Electrical Schematics ...
Page 71: ...9 6 ULT Freezer VWR International Section 9 Electrical Schematics ...
Page 72: ...ULT Freezer 9 7 VWR International Section 9 Electrical Schematics ...
Page 73: ...9 8 ULT Freezer VWR International Section 9 Electrical Schematics ...
Page 74: ...ULT Freezer 9 9 VWR International Section 9 Electrical Schematics ...
Page 75: ...9 10 ULT Freezer VWR International Section 9 Electrical Schematics ...
Page 76: ...ULT Freezer 9 11 VWR International Section 9 Electrical Schematics ...
Page 77: ...9 12 ULT Freezer VWR International Section 9 Electrical Schematics ...
Page 78: ...ULT Freezer 9 13 VWR International Section 9 Electrical Schematics ...
Page 79: ...9 14 ULT Freezer VWR International Section 9 Electrical Schematics ...
Page 80: ...ULT Freezer 9 15 VWR International Section 9 Electrical Schematics ...
Page 81: ......
Page 90: ...VWR International P O Box 5015 Bristol CT 06011 8960 ...