Certifications
Declaration of Conformity is available from the factory
Safety Specifications
Indoor Use Only
Altitude - up to 2,000 meters
Temperature - 5°C to 43°C
Humidity - Maximum RH 80% for temperatures up to 31°C, decreasing linearly to 50% RH at 40°C
Mains Supply Fluctuations - Mains supply voltage fluctuations not to exceed ±10% of the nominal volt-
age
Installation Category II
1
Pollution Degree 2
2
Class of Equipment I
______________________________________________________________________
1 Installation category (overvoltage category) defines the level of transient overvoltage which the instrument
is designed to withstand safely. It depends on the nature of the electricity supply and its overvoltage protection
means. For example, in CAT II which is the category used for instruments in installations supplied from a
supply comparable to public mains such as hospital and research laboratories and most industrial laboratories,
the expected transient overvoltage is 2500V for a 230V supply and 1500V for a 120V supply.
2 Pollution degree describes the amount of conductive pollution present in the operating environment.
Pollution degree 2 assumes that normally only non-conductive pollution such as dust occurs with the
exception of occasional conductivity caused by condensation.
6-2
ULT Freezer
VWR International
Section 6
Specifications
Model
5721
5720
Temperature Range
-50°C(-58°F) to -86°C(-123°F) in an 18C to 32C * (64.4F to 89.6F) ambient
Exterior Dimensions
96.0”W x 40.5” H x 28.8” F-B
(243.8cm x 102.9cm x 73.2cm)
Add 7.9” (20.1cm) to F-B for wall spacer/handle.
Interior Dimensions
66.5”W x 28.0”H x 18.5” F-B
(168.9cm x 71.1cm x 47.6cm)
Capacity
20.0 cu. ft. (566.3 liters)
Refrigeration
Cascade system, (2) hermetically-sealed compressors
Insulation
Non CFC, foamed-in-place urethane: 5.0" (12.7cm) cabinet; 2.0" (5.0cm) lid;
1.0" (12.5cm) sub-lids
Electrical
120VAC, 1 PH, 60 Hz, 16.0 FLA
Operating Range: 108VAC-130VAC
230VAC, 1 PH, 50/60 Hz, 12.0 FLA
Operating Range: 208VAC-240VAC
Breaker Requirements
20 Amp, 120VAC, Dedicated Circuit,
20 Amp Time Delay Breaker
15 Amp, 230VAC, Dedicated Circuit,
15 Amp Time Delay Breaker
Shipping Weight
833 lbs. ( 377.8 kg)
Summary of Contents for 5700 Series
Page 19: ......
Page 21: ......
Page 41: ......
Page 44: ...ULT Freezer 7 1 VWR International Section 7 Parts ...
Page 45: ...7 2 ULT Freezer VWR International Section 7 Parts ...
Page 46: ...ULT Freezer 7 3 VWR International Section 7 Parts ...
Page 59: ......
Page 60: ...ULT Freezer 8 1 VWR International Section 8 Refrigeration Schematics ULT Freezer 8 1 ...
Page 61: ...8 2 ULT Freezer VWR International Section 8 Refrigeration Schematics ...
Page 62: ...ULT Freezer 8 3 VWR International Section 8 Refrigeration Schematics ...
Page 63: ...8 4 ULT Freezer VWR International Section 8 Refrigeration Schematics ...
Page 64: ...ULT Freezer 8 5 VWR International Section 8 Refrigeration Schematics ...
Page 65: ......
Page 66: ...ULT Freezer 9 1 VWR International Section 9 Electrical Schematics ...
Page 67: ...9 2 ULT Freezer VWR International Section 9 Electrical Schematics ...
Page 68: ...ULT Freezer 9 3 VWR International Section 9 Electrical Schematics ...
Page 70: ...ULT Freezer 9 5 VWR International Section 9 Electrical Schematics ...
Page 71: ...9 6 ULT Freezer VWR International Section 9 Electrical Schematics ...
Page 72: ...ULT Freezer 9 7 VWR International Section 9 Electrical Schematics ...
Page 73: ...9 8 ULT Freezer VWR International Section 9 Electrical Schematics ...
Page 74: ...ULT Freezer 9 9 VWR International Section 9 Electrical Schematics ...
Page 75: ...9 10 ULT Freezer VWR International Section 9 Electrical Schematics ...
Page 76: ...ULT Freezer 9 11 VWR International Section 9 Electrical Schematics ...
Page 77: ...9 12 ULT Freezer VWR International Section 9 Electrical Schematics ...
Page 78: ...ULT Freezer 9 13 VWR International Section 9 Electrical Schematics ...
Page 79: ...9 14 ULT Freezer VWR International Section 9 Electrical Schematics ...
Page 80: ...ULT Freezer 9 15 VWR International Section 9 Electrical Schematics ...
Page 81: ......
Page 90: ...VWR International P O Box 5015 Bristol CT 06011 8960 ...