128 vwr.com Instruction manual VWR
®
Imager Chemi Premium
Warranty
VWR
warrants that this product will be free from defects in material and workmanship for a period of
two (2) years from date of delivery. If a defect is present, VWR will, at its option and cost, repair,
replace, or refund the purchase price of this product to the customer, provided it is returned during the
warranty period. This warranty does not apply if the product has been damaged by accident, abuse,
misuse, or misapplication, or from ordinary wear and tear. If the required maintenance and inspection
services are not performed according to the manuals and any local regulations, such warranty turns
invalid, except to the extent, the defect of the product is not due to such non-performance.
Items being returned must be insured by the customer against possible damage or loss. This warranty
shall be limited to the aforementioned remedies. IT IS EXPRESSLY AGREED THAT THIS
WARRANTY WILL BE IN LIEU OF ALL WARRANTIES OF FITNESS AND IN LIEU OF THE
WARRANTY OF MERCHANTABILITY.
Compliance with local laws and regulations
The customer is responsible for applying for and obtaining the necessary regulatory approvals or other
authorisations necessary to run or use the Product in its local environment. VWR will not be held
liable for any related omission or for not obtaining the required approval or authorisation, unless any
refusal is due to a defect of the product.
Imager Chemi Premium Troubleshooting
PROBLEM
POSSIBLE CAUSE
SOLUTION
Camera not connected or
not responding.
Camera USB cable not
plugged in.
Software driver for the
camera is missing.
Make sure all cables are
connected as shown in
Instrument Set-up and/or the
Installation Quick Guide.
Install camera driver.
Press the refresh button on
the Imager Chemi Premium
icon.
'Hardware not available'
message displayed.
You do not have the correct
filter or lighting for the dye
you have selected.
Check the 'hardware list' on
the
Hardware
screen.
Contact VWR
to upgrade your system.
Programme in any extra
lights or filters that you may
have installed on the
Hardware
screen.
Dye not in list.
The dye that you are using is
not currently listed in the
database.
Contact VWR
Dye name is greyed out.
You do not have the
appropriate hardware to
image the dye that you have
selected.
Check the 'hardware list' on
the
Hardware
screen.
Contact VWR
to upgrade your system.
Programme in any extra
lights or filters that you may
have installed on the
Hardware
screen.
Blue Light LED
Transilluminator will not
turn on.
Make sure the
Transilluminator switch is in
the ON position.
If the Transilluminator will not
turn on then please contact
VWR