Instruction manual |
9
5
Screw each objective into the thread of the nosepiece,
clockwise with increasing magnification. (Fig. 5)
6
Insert the power supply jack in the connector placed at the
rear side of the microscope. (Fig. 6)
7.1.1 Installing fluorescence filter
This procedure is only necessary for TL534B FL and
TL534B-SA FL4 models. TL534B FL BG model comes with B and
G cubes preinstalled.
1
Disconnect the power supply plug from the fluorescence
illuminator.
2
Open the side cover of the illuminator, by unscrewing the side
screws (1). (Fig. 7)
It could be helpful to remove the observation head.
The cubes are mounted on the opposite side of the cover:
opening the left cover acts on the right side of the slider and
vice versa.
3
Open the top door of the fluorescence illuminator by
unscrewing the four screws (2) and release the cover. (Fig. 8)
FIGURE 5
FIGURE 6
1
FIGURE 7
2
FIGURE 8