18
Time
Single injection time, range: 1 ~ 300s, interval 1s adjustment or set to "-"
continuous injection until manual stop.
Speed
Sample inhalation speed, range: 1 ~ 8ml / min, interval 0.5 ml / min
adjustment.
Recycling samples
1.
Place the discharge end of the silicone tube in the waste collection container;
2.
Main interface, press the button
Recycling
on the right, pop-up recycling control
parameter interface, press the button
<
,
>
to set the injection parameters
according to the requirements. Press the button
OK
to start recycling, reach the set
time (or press the button
Stop
) end. Press the button
Cancel
to return to the main
interface.
Time
Time for single sample recycle, range: 1 ~ 300s, interval 1s adjustment or
set to "-" continuous injection until manual stop.
Cleaning the flow cuvette
1.
Place the suction end of the silicone tube in the cleaning solution;
2.
Main interface, press the button
Cleaning
on the right to pop up the cleaning
control parameter interface, press the button
<
,
>
to set the injection parameters
according to the requirements. Press the button
OK
to start recycling, reach the set