22
Maintenance of the FG chucking system
of FG contra-angle handpieces
> Clean the chuck at least once a week.
> Remove the rotary instrument.
> Use the T1 adaptor 02693000
With its spray exit nozzle
directed downwards, push
the T1 adaptor onto the
maintenance attachment.
Initially, keep the cover open.
Push the instrument onto the
adaptor from underneath.
Support the instrument at the
edge of the cover.
Press the program button once
for about 2 seconds.
Press the instrument against
the adaptor for about
10 seconds.
Close the cover. Allow the
remainder of the maintenance
cycle to complete.
1 x
2 secs.