42
7.3
REAL-TIME CYCLE DATA INFORMATION
Press the “Info” icon at any time during the sterilization cycle or when the sterilizer is idling to view a list of
the most important technical parameters of the unit (see screen below, right).
The INFO screen is controlled by an automatic timeout. Press the “Pause” icon to prevent the screen from
automatically returning to the previous screen.
See the table below for a description of the displayed technical parameters:
Legend:
P Chamber
Pressure/vacuum inside the sterilization chamber
T Theo.
Theoretical temperature calculated from chamber pressure
T Chamber Heat
Temperature of the chamber heating element
T Steam gen.
Temperature of the steam generator heating element
T Condenser
Temperature of the heat exchanger (condenser)
Power Steam gen.
Power consumption of steam generator
Power Chamber heat.
Power consumption of the chamber jacket heater
Cond. H20
Water conductivity
H20
Water volume pumped inside the steam generator
Manual
stop
Info
Scroll-up
Scroll-down
Pause