50
11.2 CLEANING THE DOOR SEAL
ATTENTION!
Operation to be carried out when the
sterilization chamber is cold
Clean the door seal and the chamber face side with a non ab-
rasive cloth and a mild detergent solution. Rinse with clean wa-
ter.
11.3 CLEANING THE CHAMBER AND CHAMBER COMPONENTS
ATTENTION!
Operation to be carried out when the sterilization chamber is cold
•
Remove the trays from the chamber.
•
Remove the chamber rack.
•
Clean the chamber with a damp sponge and a mild detergent solution.
•
Rinse the chamber with a sponge and remove all traces of the cleaning agent.
Apply the same procedure for the rack, trays or cassettes.
•
Thoroughly clean all around the sterilization chamber and the internal chamber in-
terface at the back of the chamber.
•
Do not bend or damage the temperature sensor at the bottom/right of the chamber.
•
Never use disinfectants or sharp objects to clean the chamber.
11.4 CLEANING THE CHAMBER FILTER
ATTENTION!
Operation to be carried out
when the sterilization chamber
is cold
•
Empty the sterilization chamber by removing
the trays and rack.
•
Remove the filter cap at the back of the
chamber (bottom/center) by turning it coun-
ter-clockwise.
•
Remove the cartridge filter and rinse it with
tap water.
•
Insert the filter on the cap, plug it in and lock
it by turning clockwise.