4.7 PEAK hold
The Peak kold function captures the peak AC or DC
voltage or current. The meter can capture negative or
positive peaks as fast as 1 millisecond in duration.
1. Momentarily press the PEAK button, ‘PEAK’ and
‘MAX’ will display in left auxiliary display. MIN’ will
display in right auxiliary display. The multimeter will
update the dispay each time a lower negative peak
occurs.
2. Press the EXIT button to exit the PEAK HOLD mode.
Auto Power Off feature will be disabled auto
matically in this mode.
4.8 Data record(STORE/RECALL)
STORE
function
In the current testing mode, press STORE button one
time, enter into STORE function.
On the left upper corner of LCD shows NO
XXXX S which states current storage serial
number. On the right upper corner of LCD
shows XXXX S which states how many current
storage is used.
1. On the left upper shows 0000 S, which states
recording interval time; using & - to
select S the range is 0~255 seconds
2. When the recording interval time is 0000 S then
press STORE button again to change into manual
recording. Press the STORE button again to record
once
3. When the recording interval time is 1~255 S then
press STORE button again to start recording auto-
matically from 0000. Recording times is showed on
the left upper corner, data is showed on the right
upper corner (Due to digitally limitation, there is only
display preceding four numbers)
To finish above STORE function, press EXIT-button
shortly. If you want to clean all the memory data, the
steps are: When power on, press the RANGE-button a
long time. The LCD will flash and the buzzer will sound,
which means all memory data have been cleared.