The WAGER Model 6500 Smoke Meter has been designed to be nearly trouble
free for many years of use.
However, some components are subject to unintentional abuse, which is not
covered in our warranty.
Error Messages
CABLE FAILURE DISCONNECTED?
The light source is fully powered, but no light is detected receiver. This could be
caused by several factors:
=
The sensor head assembly cable is not connected to the control unit
=
The light source in the sensor head assembly has failed.
=
The connecting cable is malfunctioning.
Action:
Check to make sure that the cable is firmly attached to the unit.
If another unit’s sensor head assembly and cable is available, try to see if this corrects
the error.
If the message disappears, then you need to send in the sensor head assembly and/or
cable for repair.