8.3 History Mode
In History mode all data collection and transfer is controlled by the dual stage
trigger operation. In summary, the
T1
position is used to align the instrument
on the chosen target and the
T2
position is used to generate the logged data.
With the trigger pulled to the
T1
position, operation is exactly the same as in
Classic mode. All four temperature types (Peak, Continuous, Average and
Valley) are displayed with the selected (highlighted) type being displayed
internally within the viewfinder as well as being streamed serially in 0.5 second
intervals (wired connection or Bluetooth).
If the trigger is pulled to the
T2
position, this is confirmed by a single ‘blink’ on
the internal viewfinder, coupled with a single ‘beep’ on the sounder (if switched
on). When the trigger is pulled to the
T2
position, both the viewfinder and side
displays are ‘frozen’ and a ‘packet’ of data is generated and sent out for data
logging.
This packet of data comprises the overall peak, last continuous,
overall average and overall valley temperature measurements taken during
the time that the trigger was pressed to the
T1
position. The emissivity set
point and window compensation indicator value in use are also output, as part
of the data packet.
The temperature components of this data packet are also retained within the
instruments memory and may be viewed on the side display by going into
Menu mode and re-selecting twice on the History mode icon.
Note that further data packets will only be added after each press of the trigger
to the
T2
position and the data added will be the measurements made at the
T1
position during the time preceding the press to
T2
. Any measurement
made at the
T1
position without going to the
T2
position will not be stored.
If additional data is added, any higher peak (or lower valley) temperature
recorded will overwrite the earlier stored value. The new average value will
then become the average over all stored data and the continuous value
displayed will become an average of the last continuous values for each data
packet within the overall stored sets.
When the stored data set has been inspected the user has the option of deleting
the data by selecting the ‘waste basket’ icon, or exiting the data set without
deletion (exit icon) in which case further data may be added to the stored set.
Note
If the instrument is switched off,
all
stored data will be lost.
Summary of Contents for Heat Spy DHS520
Page 2: ......
Page 3: ......
Page 9: ......
Page 12: ......
Page 14: ...Page 2 Wahl DHS520 User Manual Fig 1 Wahl DHS520 520B portable infrared thermometer ...
Page 21: ......
Page 35: ...User Manual Wahl DHS520 Page 22 Blank ...
Page 37: ......
Page 38: ...Wahl DHS520 Page 24 Blank UserGuide ...
Page 39: ...User Manual Wahl DHS520 ...
Page 40: ......
Page 45: ......
Page 51: ......
Page 59: ...Page 43 User Manual Wahl DHS520 ...