255.805.001
32
Date: 2021-03-12
Checklist
Use the checklist accordingly; before each use, each month etc.
Check
Type
Result
Installed / mounted correct
Inspection
Display and LEDs visible for the operator
Inspection
Brake output
Functional
Encoder input and attachment to drive
Functional
Hard-limit inputs
Functional
Emergency stop input
Functional
powerCON TRUE1 or CEE 16A
Inspection
powerCON TRUE1
Regularly check the condition of the powerCON TRUE1 connector system (cable connector and
chassis) for clear signs of wear and tear.
Through improper use and the excessive use of force, the encoding lugs and guide slots in the
connector system can become so severely worn or damaged that it is possible to insert the cable
connector the wrong way. However, when inserted the wrong way, the cable connector should
not twist-lock into place.
A cable connector inserted the wrong way could, in some circumstances, lead to contact
between
live wires and the grounding contact in the plug socket.
To determine that the cable connector is inserted correctly, follow the use instructions and
listen for the “click” when performing the twist-lock. Immediately unplug any connector that
does not twist-lock correctly in place.