29
TROUBLESHOOTING
If any abnormality arises during use, please check the following points
Note:
If the unit still does not work, contact HoMedics Consumer Relations.
Under no circumstance should you disassemble or attempt to repair the
unit by yourself. Contact information for HoMedics Consumer Relations
Department can be found on the warranty page.
SYMPTOMS
POSSIBLE
CAUSES
CORRECTION
Unit does not turn
on when the
START/
STOP button
is
pushed.
Batteries have
run down.
Replace them with
four new AA
alkaline batteries.
Battery polarities
have been positioned
incorrectly.
Re-insert the batteries
in the correct positions.
EE measurement error
symbol shown on dis
-
play or the blood pres
-
sure value is displayed
excessively low
(or high).
Cuff has been placed
incorrectly.
Wrap the cuff properly
so that it is positioned
correctly.
Did you talk or move
during measurement?
Measure again. Keep
arm steady during
measurement.
Shaking of the arm with
the cuff on.
E1 error symbol shown
on display
Air circuit abnormality.
Cuff tube may not be
plugged into monitor
correctly.
Check cuff connection.
Measure again.
E2 error symbol shown
on display
Inflation pressure
exceeding 300 mmHg.
Switch the unit off, then
measure again.
E3 error symbol shown
on display
Error determining
measurement data.
Measure again.